--------- MULTIMETER CURRENT CLAMP
CPM
-------- PG No. 16
-
3.2.4.- Frequency measurement
1. Set the rotary switch [1] to the Hz position.
2. Connect the red test probe to the terminal [5] and the
black test probe to the COM terminal [4].
3. Connect the test probes in parallel with the points to be measured and read
the frequency value on the display [6].
3.2.5.- Continuity sound alarm
1. Set the rotary switch [1] to position“ “ .
2. Connect the black test probe to the COM terminal [4] and
the red test probe to the terminal [5].
3. Connect the test probes to the points to be verified.
4. If the resistance value is less than 30 the built in alarm will sound.