7. ERROR CODES
After switching the meter to the network, messages about errors can
appear. Reasons about errors are presented below.
The appearance of below mentionned symbols on digital displays
means:
Overflow of upper value of programmed indication
range.
Overflow of lower value of programmed indication
range.
ErCAL Loss of meter calibration values. One must contact
the service workshop.
EroUt Loss of calibration values of meter analog outputs.
The pressure of the ESC button switches the message
off, analog outputs remain switched off. One must
contact the service shop.
Er EE Innapropriate values in meter configurating data. The
pressure of the ESC button switched the message off.
One must set meter parameters again.
ErEnr Incorrect energy values in the meter. The pressure of
the ESC button switched the message off. Energies
are reset.
ErCod Password incorrectly introduced.