Circutor GETEST Series Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Circutor GETEST Series. CIRCUTOR GETEST Series User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 20
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
Instruction manual
Vista de página 0
1 2 3 4 5 6 ... 19 20

Resumo do Conteúdo

Página 2

10 Instruction manual Getest 3.- Switch on the circuit breaker (6). The red led will light. 4.- Turn the safety key to the “ENABLED” position. Th

Página 3 - Features

11 Instruction manual Getest 8.- Type the Maximum Defect Current (16) in order to calculate the step and contact voltage. Select to measure with a

Página 4

12 Instruction manual Getest Connections Ground to test Metallic parts Min. 40 m Auxiliary ground 20 m 20 m Contact voltage 1 m Min. 40 m Auxil

Página 5

13 Instruction manual Getest Ground Resistance Note about auxiliary ground: The auxiliary ground must be good enough for performing a proper meas

Página 6

14 Instruction manual Getest Connecting to a PC Using a PDA to control the MPC unit allows to store all data in the PDA memory, and transfer them t

Página 7

15 Instruction manual Getest Error and warning messages Error message Cause Solution Error:Reading cal data. Loading default data Corruption on

Página 8

16 Instruction manual Getest Danger: loop resistance too high, possible connection error. Check connectons and ensure nobody is touching the injecti

Página 9 - Operation instructions

17 Instruction manual Getest Specifications Mains supply Voltage 230V+/-20% 50-60Hz Current 15A AC Note: Required low impedance in mains supply l

Página 10

18 Instruction manual Getest ETSI EN 301 489-17 EN61000-3-2, EN61000-3-3 EMC compliancy Device is compliant with EU LVD 2006/95/EC and EMC 89/336/EE

Página 11

19 Instruction manual Getest Notes

Página 12 - Connections

2 Instruction manual Getest

Página 13

20 Instruction manual Getest CIRCUTOR S.A. Vial Sant Jordi s/n - 08232 - Viladecavalls (Barcelona) Spain Tel: (+34) 93 745 29 00 Fax: (+34) 93 745

Página 14 - Connecting to a PC

3 Instruction manual Getest Features - Step and contact meter, based on injection of current during 1 mains cycle - Supply up to 50A

Página 15 - Error and warning messages

4 Instruction manual Getest

Página 16

5 Instruction manual Getest 10 Connect 12 Current to inject 15 1K Ohm resistor 17 Start sequence 19 Results tab 11 COM port 13 Mains frequency 16 M

Página 17 - Specifications

6 Instruction manual Getest 30 Save data button

Página 18

7 Instruction manual Getest Safety Electrical Shock This device may generate voltages that may be harmful to the body or cause death. Several pro

Página 19

8 Instruction manual Getest Heavy weight Weight of the device is 45kg. This weight may be harmful to the body if not lifted by trained persons,

Página 20

9 Instruction manual Getest Operation instructions 1.- Plug measure connections (2) and current connections (3) depending on the measure that is go

Comentários a estes Manuais