DIELECTRIC STRENGTH TESTER FOR ISOLATING OILS OT2-60D --------- Page 15
- Mount the electrodes in the cell, fill it with clean and new oil, of the
same type than the one that is going to test later, and elevate the
voltage until rupture 24 times.
In the case of changing the type of the test oil, eliminate all the rests of the
previous oil using a suitable dissolvent, wash the cell with the same oil that you
are going to test, drain the cell and return to fill up it to carry out the test.
During the periods of inactivity it is recommended to keep the cell full of clean
and new oil in a clean enclosure, dry and free of dust.
9. ENVIRONMENT CONSIDERATIONS
Once finalized the tests the opportune measures will be done to manage the used oil,
since this one is an injurious substance for the environment. If you don’t know how to
deal with this type of material, contact with your manager of remainders.